Uv Vis Double Beam Spectrophotometer-LMSP-UV1900
|
|
 |
|
|
The innovative optical layout and state of art mono- chromator with high grade blazed holographic grating ensure accuracy. Its integrated design assures long term stability and durability
|
|
|
The precisely aligned detector and quality deuterium and halogen lamps enhance the precision across the UV / Vis spectrum starting from 190 nm and into the near-infrared 1100nm
|
|
|
The two detectors are used to measure sample and reference respectively and simultaneously for optimizing measurement accuracy
|
|
|
Lamp Selection enables conserving the life of D2 & Tungsten lamps. Real time clock for date & time stamping of results
|
|
|
Big LCD display. (320 x 240 Dots Matrix)
|
|
|
Basic Mode, Quantitative, Kinetics, Multi - Wavelength, Wavelength Scan, DNA / Protein Test &Utility
|
|
|
|
FUNCTIONS OF SOFTWARE
|
|
BASIC MODE
|
| To measure the Absorbance and Transmittance |
| |
|
QUANTITATIVE
|
|
1. Coefficient Method
|
|
2. Standard Method curve.
|
|
Up to 10 standard samples may be used to establish a curve. Four methods to fitting a curve through the calibration points: Linear fit through zero, square fit & cubic fit.
|
| |
|
WAVELENGTH SCAN
|
|
1. The wavelength scan intervals are 0.1, 0.2, 0.5, 1, 2 & 5 nm
|
|
2. Fast, Medium & Slow Scan speeds are available. They very from100 to 1800 nm/min
|
|
3. Wavelength are scanned from high to low so that the instrument waits at high Wavelength and minimizes the degradation of UV sensitive samples
|
| |
|
MULTI-WAVELENGTH
|
|
Up to 32 wavelengths can be selected and multiple samples can be measured
|
|
|
|
KINETICS
|
|
This mode may be used for time course scanning or reaction rate calculations. Abs. vs. time graphs is displayed on the screen in real time
|
|
Wait time and measurement time up to 12 hours may be entered with time intervals of 0.5,1,2,5,10,30 seconds & 1min. Post- run manipulation includes re-scaling, curve tracking and selection of the part of the curve required for the rate calculation. Rate is calculated using a linear regression algorithm before multiplying by the entered factor
|
| |
|
DNA/PROTEIN TEST
|
| Concentration and DNA purity are quickly and easily calculated: Absorbance ratio s 260nm/280nm with optional subtracted absorbance at 320nm. |
|
DNA concentration= 62.9x A260-36.0x A280
|
|
Protein concentration=1552xA260-757.3xA280
|
| |
|
UTILITY
|
|
Has following functions
|
|
WL Reset, Printer, Lamp, Clock, Dark Current, Accu. Validity, WL Validity Connect to PC, Beeper On/Off & System Base Line
|
|
|
| |
| Microprocessor Double Beam |
|
UV Visible Spectrophotometer
|
|
| |
|
|
 |
|
Model
|
LMSP-UV1900
|
LMSP-UV1900A
|
LMSP-UV1900S
|
LMSP-UV1900AS
|
|
Make
|
LABMAN
|
|
Cell Holder
|
10mm single cuvette Holder
|
10mm 8cell Programable Holder
|
10mm single cuvette Holde
|
10mm 8 cell Programable Holder
|
|
Bandwidth
|
1nm
|
0.5 /1.0/2.0/4.0 nm Variable
|
|
Optical System
|
Double Beam, Grating 1200 lines / mm
|
|
Wavelength Range
|
190 - 1100nm
|
|
Wavelength Accuracy
|
+ 0.3nm
|
|
Wavelength Repeatability
|
+ 0.2nm
|
|
Scanning Speed
|
Fast, Medium & Low
|
|
Photometric Accuracy
|
+0.3%T
|
|
Photometric Repeatability
|
+ 0.2%T
|
|
Photometric Range
|
-0.3-3A,0-200%T,0-9999C
|
|
Stability
|
+ 0.001A/h @ 500nm
|
|
Baseline Flatness & Noise
|
+0.001A
|
|
Stray Light
|
<0.05%T @ 220nm,360nm
|
|
Data Output Port
|
USB
|
|
Printer Port
|
Parallel Port
|
|
Display
|
320 x 240 Graphic LCD
|
|
Lamps
|
Deuterium Lamp & Tungsten Halogen Lamp
|
|
Detector
|
Silicon Photodiode
|
|
Power Supply
|
AC 220V/50Hz
|
|
Dimension
|
625 x 430 x 206 mm
|
|
Weight
|
30 kg
|
|
|
INSTRUMENT INCLUDES: 4 Nos 10mm Glass cuvettes, one pair of quartz cuvette, Manual & Dust Covers.
|